Meteorological environment monitoring equipment supplier
Insist on doing high-precision customer favorite technology products
1、Rainwater and sewage drainage network water quality monitor Product Overview
Rainwater and sewage drainage network water quality monitor can monitor the water quality of the Rainwater and sewage networks in polluting enterprises, industrial parks and towns.Rainwater and sewage drainage network water quality monitorX transmits data directly to a designated server via 4G, supporting the MQTT protocol for transparent transmission of JSON data.
Data in other formats is transmitted via 4G to our cloud platform.We provide an HTTP interface for other platforms to access data.
The built-in 8GB TF card allows data to be stored in the memory card.
The baud rate of each serial port can be modified via commands, and required sensor parameters can also be added, modified, and deleted via commands.
2、Rainwater and sewage drainage network water quality monitor Features
1.Built-in 3.6V 152AH lithium battery, guaranteed for three years of continuous operation.
2.IP68 waterproof, no worries about pipe flooding.
3.Two 485 interfaces: one for device configuration and one for sensor reading.
4.Supports Modbus RTU sensor reading and data storage.
5.Serial port baud rate can be independently modified.
6.Sensor communication port supports up to 32 nodes.
3、Rainwater and sewage drainage network water quality monitor Specifications
Operating Voltage: 3.6V Internal Lithium Battery
Dimensions: 204x82x162mm
Environmental: Temperature: -40-85°C, Humidity: 5%-95%
Communication Type: RS485
Communication Protocol: Modbus RTU
Communication Distance: 300 meters
Baud Rate: 1200-115200 bps, default 9600 8 N 1
Protection Level: 600W lightning surge protection per RS-485 line, ±15kV ESD protection
4、Rainwater and sewage drainage network water quality monitor Sensor parameters
Serial number | Sensor type | Measurement range | Measurement principle | Measurement accuracy | Configuration |
1 | Temperature | 0~50℃ | High-precision digital sensor | ±0.3℃ | √ |
2 | Electrical conductivity | 0~5000uS/cm, 0~10000uS/cm | Contact method of electrode | ±1.5% | √ |
3 | pH | 0~14 (ph) | Electrochemistry (salt bridge) | ±0.1PH | √ |
4 | Dissolved oxygen | 0~20mg/L | Fluorescence lifetime method | ±2% | √ |
5 | Ammonium nitrogen | 0-1000.00mg/L(default) 0-100.00mg/l/ (customizable) | Ion selective electrode method | 10% of the reading, ±0.5℃ | √ |
6 | Turbidity | 0~40NTU (optional) | Scattering light method | ±1% | √ |
7 | Cash on delivery | 0~500mg/L | UV254 absorption method | ±5% |