Meteorological environment monitoring equipment supplier
Insist on doing high-precision customer favorite technology products

1、Dissolved Ozone Sensor Product Overview
Dissolved Ozone Sensor utilizes a dual-platinum constant-voltage measurement principle; suitable for measuring ozone levels in water, easy to install, and features low maintenance costs.Dissolved Ozone Sensor is a three-electrode potentiostatic sensor developed by our company, featuring proprietary intellectual property rights. It is designed to measure the concentration of disinfectants such as ozone and is primarily used in applications including treated water at water treatment plants, water distribution networks, secondary water supply systems, end-of-line water supply points, and swimming pools. The sensor incorporates an ARM processor and high-efficiency filtering algorithms to effectively mitigate noise interference. It is equipped with an RS-485 interface, facilitating seamless integration into computer-based or networked monitoring systems.
2、Dissolved Ozone Sensor Measurement Principle
The constant-potential sensor employs a three-electrode measurement system consisting of two platinum electrodes and one silver/silver chloride electrode. It incorporates a high-precision potentiostat that maintains a stable potential at the working electrode. Disinfectant components—such as hypochlorous acid—undergo redox reactions at the working electrode, generating a current that follows Faraday’s law, which allows for the determination of the disinfectant concentration.
3、Dissolved Ozone Sensor Product Features
1. The electrochemical sensor consumes no reagents and produces no pollutant emissions.
2. The membrane-free design eliminates the need for membrane replacement or the addition of electrolyte.
3. A three-electrode configuration ensures a stable zero point and high sensitivity.
4. An integrated high-precision sampling circuit provides excellent linearity.
5. When pH levels remain relatively stable, the sensor’s readings can be accurately compared with those obtained via the DPD measurement method.
4、Dissolved Ozone Sensor Technical Specifications
The technical specifications of the product are shown in Table 4.1.
| System | Power supply | DC 9-28V |
| Power | 0.5W (peak power) | |
| Measurement parameters | Ozone, temperature | |
| Communication Interface | RS485+4-20mA | |
| Electrode material | Platinum, silver chloride | |
| Design life | >18 months | |
| Water sample flow rate | >400mL/min, and the flow rate must be kept stable | |
| Operating temperature | -5-45℃ (non-freezing) | |
| Withstand pressure | <0.1MPa | |
| Ozone | Measurement Methods | Three-electrode amperometric method |
| Range | 0-5/20mg/L | |
| Resolution | 0.001mg/L | |
| Detection Limit | 0.030mg/L or less | |
| Accuracy | ±3% (DPD comparison error ±10% or ±0.05 mg/L, whichever is greater) | |
| Repetitive | ±3% | |
| Response Time (T90) | ≤90 seconds | |
| Calibration Methods | Single-point calibration, two-point calibration | |
| Sample pH range | 4-9 | |
| Temperature | Measurement Methods | Thermistor method |
| Range | -5-60℃ | |
| Resolution | 0.1℃ | |
| Accuracy | ±0.4℃ | |
| Repetitive | 0.2℃ | |
| Response Time (T90) | ≤25 seconds | |
| Calibration Methods | Single-point calibration |
Dissolved Ozone Sensor Product Dimensions Drawing